Table 1: Overview of The Spaulding Classification.1-6
Spaulding Classification | Medical Device Contacts | Risk of Infection Transmission | Disinfection Level |
Critical | Sterile tissue or the bloodstream | High | Sterilisation* |
Semi-critical | Mucous membranes or non-intact skin | Medium | High Level Disinfection (HLD) |
Non-critical | Intact skin only | Low | Intermediate level (ILD) or Low level disinfection (LLD) |
* Critical ultrasound probes can be high-level disinfected and used with a sterile sheath if sterilisation is not possible.4,5 In Scotland and Ireland, critical probes to guide needles or scan wounds are considered semi-critical.2,3
HLD must be used for critical ultrasound probes if sterilisation is not possible, a sterile sheath must also be applied to the probe.
Ultrasound probes are delicate, heat sensitive medical devices that cannot undergo steam sterilisation. Low temperature sterilisation methods require the probe to be out of circulation for extended periods of time, disrupting clinical workflow.
Critical probes can be high level disinfected if sterilisation is not possible, however the probe must also be covered with a sterile sheath at use.
"All critical probes (probes contacting sterile tissues or blood) should be preferably sterilised, but if sterilisation is not possible, they should be minimally high-level disinfected and used with a sterile sheath.”
"Consequently any US-assisted invasive procedure or any procedure potentially causing micro-trauma to the skin or mucous membranes has to be categorised as critical."

HLD is minimally required for semi-critical ultrasound probes, a sheath must also be applied to the probe.
“Semi-critical probes contact mucous membranes and noninvasive probes contacting non-intact/broken skin should be high-level disinfected either manually or with automated systems even if a sheath is used.”
"Under the Spaulding Classification Ultrasound Probes that come into contact with broken skin or intact mucous membranes are considered semi‐critical devices and should undergo manual cleaning followed by High Level Disinfection (HLD) between patient use."
"However the advancement of ultrasound scanning means they are increasingly used for scanning the skin which can be broken through the insertion of vascular devices or for the assessment of complex wounds.This increases the risk of contamination of the probe with blood which requires a high level disinfection process."

HLD can optionally be used for non-critical probes, however low level disinfection is sufficient.
Non-critical probes will only contact healthy, intact skin. Low level disinfection (LLD) is sufficient for these probes. However, optionally performing HLD ensures the probe is ready for any subsequent procedure.
Learn about HLD with the trophon®2 device. The trophon2 device can integrate with the range of clinical workflows to help keep your patients safe from cross-infection.
Over 1,000 probes from 25 manufacturers are tested, approved and endorsed for use with trophon devices. View the probe compatibility list.
Read about guidelines on reprocessing semi-critical and critical ultrasound probes.


The trophon® family includes trophon® EPR, trophon®2 and trophon®3 which share the same core technology of 'sonically activated' hydrogen peroxide.